Difference between revisions of "Soil moisture - capacitance (5TE Decagon)"

From Experimental Hydrology Wiki
Jump to navigation Jump to search
Line 30: Line 30:
 
==Links==
 
==Links==
 
Projects that used the above equipment:
 
Projects that used the above equipment:
 
+
* [http://teodoor.icg.kfa-juelich.de/overview-de TERENO]
 
Other related web sites:
 
Other related web sites:
 
*[http://www.decagon.com/assets/Uploads/Calibration-Equations-for-the-ECH2O-EC-5-ECH2o-TE-and-5TE-Sensors.pdf Calibration Equations or the ECH2O EC-5, ECH2O-TE and 5TE Sensors]
 
*[http://www.decagon.com/assets/Uploads/Calibration-Equations-for-the-ECH2O-EC-5-ECH2o-TE-and-5TE-Sensors.pdf Calibration Equations or the ECH2O EC-5, ECH2O-TE and 5TE Sensors]

Revision as of 12:42, 21 July 2011

Decagon 5 TM.jpg

Parameter to be measured:

Soil moisture, soil temperature, EC

Method:

Capacitance

Equipment:

  • 5TE Decagon
Effective range of temperature: -40°C - +50°C
Effective range of soil moisture: 0 - 60 % volumetric water content
Accuracy: ±3% without calibration, ±1% - ±2% with calibration
Dimensions: 145mm x 33mm x7 mm, three stripes

Advantages:

  • Relatively cheap
  • Easy to install
  • Compatible with many loggers

Disadvantages:

  • Calibration needed
  • Small measurement volume compared to other sensors
  • not for portable usage designed

What to watch out for:

  • Careful installation, air gaps have a sustainable influence on the measurement compared to other sensors because of the small measurement volume
  • Probe is tenderly and will break, if it is pushed in resistive media.

Problems/Questions:

Links

Projects that used the above equipment:

Other related web sites:

References

  • Rosenbaum, U., Huisman, J.A., Weuthen, A., Vereecken, H., Bogena, H.R. (2010): Sensor-to-Sensor Variability of the ECH2O EC-5, TE, and 5TE Sensors in Dielectric Liquids. VADOSE ZONE JOURNAL, 9(1), 181-186.