Difference between revisions of "Discharge - Current meter Ott C2"

From Experimental Hydrology Wiki
Jump to navigation Jump to search
Line 1: Line 1:
 
{| align="right"
 
{| align="right"
|[[Image:C2.JPG|right|400px|ADC]]
+
|[[Image:C2.jpg|right|400px|Ott C2]]
 
|}
 
|}
  
Line 17: Line 17:
  
 
==Advantages:==
 
==Advantages:==
 +
* the quasi-standard; benchmark for other methods
 
* simple no-fuss mechanics
 
* simple no-fuss mechanics
 
* "sensor" (propeller) response is directly visual
 
* "sensor" (propeller) response is directly visual
 +
* insensitive to suspended sediment, temperature, conductivity
 
* comparatively small case including wading rods
 
* comparatively small case including wading rods
  
Line 25: Line 27:
 
* poor response in low velocities
 
* poor response in low velocities
 
* maintainance of ball bearings (oil)
 
* maintainance of ball bearings (oil)
* mechanical vulnarability
+
* mechanical vulnerability
* no immediate calculation of discharge, no data storage?
+
* no immediate calculation of discharge, no data storage
  
  

Revision as of 13:51, 8 December 2014

Ott C2

Parameter to be measured:

Flow velocity (discharge)

Method:

propeller

Equipment:

C2:

  • propellers
  • axle
  • wading rod

counting unit (e.g. Ott Z400)

Advantages:

  • the quasi-standard; benchmark for other methods
  • simple no-fuss mechanics
  • "sensor" (propeller) response is directly visual
  • insensitive to suspended sediment, temperature, conductivity
  • comparatively small case including wading rods


Disadvantages:

  • poor response in low velocities
  • maintainance of ball bearings (oil)
  • mechanical vulnerability
  • no immediate calculation of discharge, no data storage


What to watch out for:

  • submerged weeds, low velocities, piranhas

Price:

  • ~ 8000 € (2008)

Links

Discharge - comparison of measurements (C2, Flowtracker, Flowsens, ADC, tracer, dipping rod)

Other related web sites:

Projects that used the above equipment: http://brandenburg.geoecology.uni-potsdam.de/projekte/sesam/


References

http://www.ott-hydrometry.de/web/ott_de.nsf/id/pa_c2_e.html