==Parameter to be measured:==
 
==Parameter to be measured:==
 
tracer concentration (, discharge)
 
tracer concentration (, discharge)
Line 26: Line 30:
 
* small internal memory  
 
* small internal memory  
 
* awkward software in vintage look and feel
 
* awkward software in vintage look and feel
* needs RS 232 for data download
+
* needs RS 232 (COM-Port) for data download
 
 
  
 
==What to watch out for:==
 
==What to watch out for:==
 +
* signal intensity over 2394 (corresponding to concentrations above 310,5 µg/L in de-ionized water) do not yield a linear response any more (oversaturation). Concentrations in de-ionized water above 333,5 µg/L are displayed and logged as 2464.  However, these detection limits seem to be lower by factor 3 in natural water (J.B.)
 
* memory runs as ring buffer and overwrites without asking
 
* memory runs as ring buffer and overwrites without asking
 
* included pipette works with two "pressure points"
 
* included pipette works with two "pressure points"
 +
* when erasing of logger memory fails or logger readout takes hours, replacing the internal Li-battery and RAM chip may help (soldering skills required)
 
* avoid uninformed spectators when injecting the tracer, it usually looks quite nasty
 
* avoid uninformed spectators when injecting the tracer, it usually looks quite nasty
  
Line 37: Line 42:
  
 
==Price:==
 
==Price:==
~ 24000 € (2009)
+
~ 24000 € (2009, Germany)
  
 
==Links==
 
==Links==

Latest revision as of 11:08, 16 December 2013

ADC

Parameter to be measured:

tracer concentration (, discharge)

Method:

in-situ fluorometry

Equipment:

  • main device (built-in battery and logger)
  • 2 optical probes with 20-m-fibre cables
  • precision pipette
  • auxiliary bottles and ropes
  • power supply
  • data cable and software

Advantages:

  • no water samples to take, complete data collection on-site
  • convenient menu-guided in-situ calibration
  • immediate computation of discharge
  • can be equipped with alternative filters (not included) to allow detection of other tracers
  • also available with two additional conductivity probes (MKT-2)
  • sturdy metal transport cases

Disadvantages:

  • bulky, minimum setup comprises two large cases
  • incomplete manual (How can the clock be set? What exactly does "RAM for 10000 measurements" mean?)
  • small internal memory
  • awkward software in vintage look and feel
  • needs RS 232 (COM-Port) for data download

What to watch out for:

  • signal intensity over 2394 (corresponding to concentrations above 310,5 µg/L in de-ionized water) do not yield a linear response any more (oversaturation). Concentrations in de-ionized water above 333,5 µg/L are displayed and logged as 2464. However, these detection limits seem to be lower by factor 3 in natural water (J.B.)
  • memory runs as ring buffer and overwrites without asking
  • included pipette works with two "pressure points"
  • when erasing of logger memory fails or logger readout takes hours, replacing the internal Li-battery and RAM chip may help (soldering skills required)
  • avoid uninformed spectators when injecting the tracer, it usually looks quite nasty

Problems/Questions:

Price:

~ 24000 € (2009, Germany)

Links

Product website

Projects that used the above equipment:


Other related web sites: Discharge - comparison of measurements (C2, Flowtracker, Flowsens, ADC, tracer, dipping rod)

References