Difference between revisions of "Water level - capacitive (Odyssey)"

From Experimental Hydrology Wiki
Jump to navigation Jump to search
Line 18: Line 18:
 
*The probe fits exactly on a 32 mm (ID) PVC pipe  
 
*The probe fits exactly on a 32 mm (ID) PVC pipe  
 
*You can replace the batteries yourself
 
*You can replace the batteries yourself
*Log-intervale can be event-controlled
+
*Log-interval can be event-controlled
 +
*Very good support from the company (prompt answering emails).
  
 
==Disadvantages:==
 
==Disadvantages:==

Revision as of 07:56, 16 July 2010

Odyssey.jpg

Parameter to be measured:

Water level

Method:

Capacitive

Equipment:

Odyssey

Advantages:

  • Cheap
  • Logging of 64k of data
  • Logger start by starting start time
  • The lack of an outer shroud makes it easy to transport the probes
  • Available in different lengths
  • Available with 2 sizes of brass weights
  • The probe fits exactly on a 32 mm (ID) PVC pipe
  • You can replace the batteries yourself
  • Log-interval can be event-controlled
  • Very good support from the company (prompt answering emails).

Disadvantages:

  • Each sensor needs to be calibrated.
  • the sensor is not compensated for temperature or electrical conductivity.
  • manufacture state the accuracy to be -+ 5mm, but in practice it tends to be -+1cm.
  • temperature dependency: "tests in water temperatures of 0°C - 20°C have shown a variation (under estimation) of approximately 5mm" (manufacture statement).
  • electrical conductivity dependency: "If the conductivity of the water is above 5000µS/cm the sensor is largely unaffected by fluctuations in conductivity whereby large changes result in very small variations in the recorded data" (manufacture statement).

What to watch out for:

  • Don’t damage the wire.

Problems/Questions:

  • After 2 years of use, two of my loggers would no longer connect to the computer. All others still worked fined.
  • One of my loggers would only record data for a few seconds. Restarting the logger and letting it run until the storage was full seemed to take care of this.
  • How do temperature/electrical conductivity influence the measurement.

Links

Projects that used the above equipment:


Other related web sites:

http://www.odysseydatarecording.com/odyssey_productsview.php?key=1

References