Soil moisture/Electric Conductivity - FDR (WET Sensor)
Jump to navigation
Jump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.
Parameter to be measured:
- Soil moisture, permittivity
- Indirect measurement of electric conductivity of pore water
- Temperature
Method:
Frequency Domain Sensor
Equipment:
- Delta-T Device: WET Sensor with 3 rods of 600 mm length, HH2 display for manual reading or a data logger
Advantages:
- easy and fast measurement of punctual soil moisture and electric conductivity of soil pore water
- manual as well as continuously logged measurements possible
Disadvantages:
- Each sensor has to be calibrated
- careful application of probe in stony soils
- Underestimating of permitivity in media with low permitivity (ε > 40, clay and organic soils)
- unaccurate values in saturated media
What to watch out for:
- Good contact between probe and medium has to be proved
Problems/Questions:
Links
Other related web sites:
References
- Balendonck, J., Bruins, M.A., Wattimena, M.R., Voogt, W., Huys, A. (2005): WET-sensor pore water EC calibration for three horticultural soils, Acta Horticulturae 691, ISSN 0567-7572, 789 - 796.
- Bouskila, F.,Persson, M., Berndtsson, R., Bahri, A. (2008), Soil water content and salinity determination using different dielectric methods in saline gypsiferous soil, HYDROLOG SCI J 53(1), 253-265.
- Hamed, Y.,Samy, G., Persson, M. (2006), Evaluation of the WET sensor compared to time domain reflectometry, HYDROLOG SCI J 51(4),671-681.
- Inoue, M., Ahmed, B.A., Saito, T., Irshad, M., Uzoma, K.C. (2008): Comparison of three dielectric moisture sensors for measurement of water in saline sandy soil, Soil Use and Management 24, 156–162, DOI: 10.1111/j.1475-2743.2008.00154.x.
- Kargas, G., Kerkides, P*, Seyfried, M, Sgoumbopoulou, A (2011): WET Sensor Performance in Organic and Inorganic Media with Heterogeneous Moisture Distribution, Soil Science Society of America Journal 75(4), 1244-1252, DOI: 10.2136/sssaj2010.0238.
- Regalado, C.M., Ritter, A., Rodriguez-Gonzalez, R.M. (2007), Performance of the commercial WET capacitance sensor as compared with time domain reflectometry in volcanic soils, VADOSE ZONE J 6(3), 1539-1663.
- Wu, J., Li, P., Qian, H., Fang, Y. (2014), Assessment of soil salinization based on a low-cost method and its influencing factors in a semi-arid agricultural area, northwest China, Environ Earth Sci., 71, 3465–3475, DOI: 10.1007/s12665-013-2736-x.